Hangzhou Biomedical Technology Co., LTD
|
Place of Origin: | Zhejiang, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
General instruction:
Portable Patient Monitor is adaptable to adult, pediatric and neonatal usage. It can monitor vital signals as ECG, Respiratory Rate, SpO2, NIBP, TEMP and IBP. It integrates parameter measuring modules, display and recorder in one device, featuring in compactness, lightweight and portability. Replaceable built-in battery facilitates transportation of patient.
Specifications
ECG
Lead mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection I, II, III, avR, avL, avF, V,
Waveform 2 ch
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III,
Waveform 1 ch
Gain: X2.5mm/mV, X5.0mm/mV, X10mm/mV, X20mm/mV, auto
Resolution 1 bpm
Sensitivity > 200 (uV P-P)
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential 300mV
Leakage Current < 10 uA
Baseline Recovery < 3 S After Defi.
ECG Signal Range: 8 m V ( Vp-p )
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
5%±Calibration Signal 1 (mV p-p), Accuracy :
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R
ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm Available
Review Available
RESP
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 ~ 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy 2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Method Oscillometric